Health & Environmental Research Online (HERO)


Print Feedback Export to File
2570228 
Journal Article 
Development of SRM 2907 trace terrorist explosives simulants for the detection of Semtex and triacetone triperoxide 
Maccrehan, W; Moore, S; Hancock, D 
2011 
Yes 
Analytical Chemistry
ISSN: 0003-2700
EISSN: 1520-6882 
83 
23 
9054-9059 
English 
Effective and accurate detection of trace explosives is crucial in the effort to thwart terrorist explosives attacks. A National Institute of Standards and Technology (NIST) standard reference material (SRM) has been developed for the evaluation of trace explosives detectors that sample by collection of residue particles using swiping or air filtration. SRM 2907 Trace Terrorist Explosives Simulants consists of two materials individually simulating the residues of the plastic explosive Semtex [for pentaerytritol tetranitrate (PETN)] and the improvised explosive triacetone triperoxide (TATP). Unique challenges were encountered in the development of these materials, including the selection of suitable inert substrates, material preparation, thermal stability testing, and analytical method development. Two independent analytical methods based on liquid chromatography with ultraviolet absorbance and mass spectrometric detection, LC-UV and LC/MS, respectively, were developed and used to certify the mass fractions of PETN and TATP. These materials were further evaluated for their suitability on a field swipe-sampled trace explosives detectors based on ion mobility spectrometry (IMS).