Jump to main content
US EPA
United States Environmental Protection Agency
Search
Search
Main menu
Environmental Topics
Laws & Regulations
About EPA
Health & Environmental Research Online (HERO)
Contact Us
Print
Feedback
Export to File
Search:
This record has one attached file:
Add More Files
Attach File(s):
Display Name for File*:
Save
Citation
Tags
HERO ID
3153588
Reference Type
Journal Article
Title
Suppression of skeletal muscle signal using a crusher coil: A human cardiac (31) p-MR spectroscopy study at 7 tesla
Author(s)
Schaller, B; Clarke, WT; Neubauer, S; Robson, MD; Rodgers, CT
Year
2016
Is Peer Reviewed?
1
Journal
Magnetic Resonance in Medicine
ISSN:
0740-3194
EISSN:
1522-2594
Volume
75
Issue
3
Page Numbers
962-972
Language
English
PMID
25924813
DOI
10.1002/mrm.25755
Abstract
PURPOSE:
The translation of sophisticated phosphorus MR spectroscopy ((31) P-MRS) protocols to 7 Tesla (T) is particularly challenged by the issue of radiofrequency (RF) heating. Legal limits on RF heating make it hard to reliably suppress signals from skeletal muscle that can contaminate human cardiac (31) P spectra at 7T. We introduce the first surface-spoiling crusher coil for human cardiac (31) P-MRS at 7T.
METHODS:
A planar crusher coil design was optimized with simulations and its performance was validated in phantoms. Crusher gradient pulses (100 μs) were then applied during human cardiac (31) P-MRS at 7T.
RESULTS:
In a phantom, residual signals were 50 ± 10% with BISTRO (B1 -insensitive train to obliterate signal), and 34 ± 8% with the crusher coil. In vivo, residual signals in skeletal muscle were 49 ± 4% using BISTRO, and 24 ± 5% using the crusher coil. Meanwhile, in the interventricular septum, spectral quality and metabolite quantification did not differ significantly between BISTRO (phosphocreatine/adenosine triphosphate [PCr/ATP] = 2.1 ± 0.4) and the crusher coil (PCr/ATP = 1.8 ± 0.4). However, the specific absorption rate (SAR) decreased from 96 ± 1% of the limit (BISTRO) to 16 ± 1% (crusher coil).
CONCLUSION:
A crusher coil is an SAR-efficient alternative for selectively suppressing skeletal muscle during cardiac (31) P-MRS at 7T. A crusher coil allows the use of sequence modules that would have been SAR-prohibitive, without compromising skeletal muscle suppression. Magn Reson Med 75:962-972, 2016. © 2015 The Authors. Magnetic Resonance in Medicine Published by Wiley Periodicals, Inc. on behalf of International Society of Medicine in Resonance.