Health & Environmental Research Online (HERO)


Print Feedback Export to File
6729438 
Journal Article 
MRI-compatible Nb-60Ta-2Zr alloy used for vascular stents: haemocompatibility and its correlation with protein adsorption 
Li, XM; Li, HZ; Wang, SP; Huang, HM; Huang, HH; Ai, HJ; Xu, J 
2014 
Materials Science and Engineering C: Materials for Biological Applications
ISSN: 0928-4931
EISSN: 1873-0191 
42 
385-395 
English 
Nb-60Ta-2Zr is a newly developed MRI-compatible alloy used for vascular stents. In this work, its haemocompatibility was investigated, including platelet adhesion (lactate dehydrogenase activity), platelet activation (P-selectin expression), coagulation and haemolysis. For comparison, parallel assessments for these factors were performed for the niobium, tantalum, 316L stainless steel (316L SS) and L605 Co-Cr alloy (L605). In addition, albumin and fibrinogen were selected to examine the correlation of protein adsorption with platelet adhesion and metal surface properties. The propensity for platelet adhesion and activation on the Nb-60Ta-2Zr alloy was at nearly the same level as that for Nb and Ta but was slightly less than those of 316L SS and L605. The mitigated platelet adhesion and activation of the Nb-60Ta-2Zr alloy is associated with its decreased adsorption of fibrinogen. The Nb-60Ta-2Zr alloy has a longer clotting time and exhibits significantly superior thromboresistance than 316L SS and L605. Moreover, the haemolysis rate of the Nb-60Ta-2Zr alloy satisfies the bio-safety requirement of the ISO 10993-4 standard. The favourable haemocompatiblity of the Nb-60Ta-2Zr alloy provides evidence of its good biocompatibility and of its suitability as a candidate stent material.