Health & Environmental Research Online (HERO)


Print Feedback Export to File
7217059 
Journal Article 
A new AC-coupled amplifier for portable ECG without reference electrode 
Wang, CS; , 
2013 
Yes 
Computers & Electrical Engineering
ISSN: 0045-7906 
PERGAMON-ELSEVIER SCIENCE LTD 
OXFORD 
141-149 
This paper proposes a single power supply operated front-end AC coupling biopotential amplifier without reference electrode for portable electrocardiogram (ECG) with dual mode input for common-mode signal and common mode DC, which can effectively improve the Common Mode Rejection Ratio (CMRR) and the common mode impedance operating on a single power supply and therefore simplifying the circuit design. The proposed circuit design has a gain of 60 dB and a single supply of +5 V achieves a CMRR over 130 dB at 60 Hz without any trimming. Actually, in comparison between ECG without Driven Right Leg (DRL) circuit and ECG with DRL circuit, the former has a built in lead-1 ECG for better signal to noise ratio. This result is also valid for portable ECG measurement. (C) 2012 Elsevier Ltd. All rights reserved.