Jump to main content
US EPA
United States Environmental Protection Agency
Search
Search
Main menu
Environmental Topics
Laws & Regulations
About EPA
Health & Environmental Research Online (HERO)
Contact Us
Print
Feedback
Export to File
Search:
This record has one attached file:
Add More Files
Attach File(s):
Display Name for File*:
Save
Citation
Tags
HERO ID
7227874
Reference Type
Journal Article
Title
Riata lead failure with normal electrical lead parameters and normal fluoroscopic appearance
Author(s)
Lakshmanadoss, U; Lahoda, D; Deshmukh, P; ,
Year
2013
Is Peer Reviewed?
Yes
Journal
Journal of Interventional Cardiac Electrophysiology
ISSN:
1383-875X
EISSN:
1572-8595
Language
English
PMID
23080330
DOI
10.1007/s10840-012-9739-5
Abstract
Recently, insulation breach of the Riata lead raised a big concern. Management of these externalized leads has been addressed by professional organizations. However, what to do in patients with Riata leads without manifested failure is an ongoing clinical dilemma. Here, we present two clinical scenarios where the implantable cardioverter defibrillator system failed to deliver shock therapy in spite of having a new ICD generator and "appropriately functioning leads" as revealed by lead interrogation.