A gas chromatography/electron capture detector (GC/ECD) technique has been developed to determine elemental iodine In toluene and cyclohexane solvents. The retention index based on the alkyI Iodide and n -paraffin series was determined. Sufficient resolution was obtained to resolve I2 from alkyl iodides. A detection limit of 39 ng of I2 was obtained. A dissociative electron capture mechanism was proposed for the ECD response to I2- Radiotracer studies established the I2 was transmitted through the column without conversion to an organic Iodide. The utility of the developed technique was demonstrated on a Agl in NH4OH mass spectrometry standard solution. An accuracy of 23% and a precision of 12% were demonstrated. 1984, American Chemical Society. All rights reserved.