Health & Environmental Research Online (HERO)


Print Feedback Export to File
1049782 
Journal Article 
Salt tolerance of desorption electrospray ionization (DESI) 
Jackson, AU; Talaty, N; Cooks, RG; Van Berkel, GJ 
2007 
Journal of the American Society for Mass Spectrometry
ISSN: 1044-0305
EISSN: 1879-1123 
18 
12 
2218-2225 
English 
The salt tolerance of desorption electrospray ionization (DESI) was systematically investigated by examining three different drug mixtures in the presence of 0, 0.2, 2, 5, 10, and 20% NaCl:KCl (1:1) from different surfaces. At physiological salt concentrations, the individual drugs in each mixture were observed in each experiment. Even at salt concentrations significantly above physiological levels, particular surfaces were effective in providing spectra that allowed the ready identification of the compounds of interest in low nanogram amounts. Salt adducts, which are observed even in the absence of added salt, could be eliminated by adding 0.1% 7 M ammonium acetate to the standard methanol:water (1:1) spray solvent. Comparison of the salt tolerance of DESI with that of electrospray ionization (ESI) demonstrated better signal/noise characteristics for DESI. The already high salt tolerance of DESI can be optimized further by appropriate choices of surface and spray solution. 
IRIS
• Methanol (Non-Cancer)
     Search 2012
          WOS