Determination of Tricyclazole in Water Using Solid Phase Extraction and Liquid Chromatography 
Nhung, DTT; Phong, TK; Watanabe, H 
2009 
Yes 
Journal of Liquid Chromatography and Related Technologies
ISSN: 1082-6076
EISSN: 1520-572X 
32 
18 
2712-2720 
English 
A method for determination of tricyclazole in water using solid phase extraction and high performance liquid chromatography (HPLC) with UV detection at 230nm and a mobile phase of acetonitrile:water (20:80, v/v) was developed. A performance comparison between two types of solid phase sorbents, the C18 sorbent of Supelclean ENVI-18 cartridge and the styrene-divinyl benzene copolymer sorbent of Sep-Pak PS2-Plus cartridge was conducted. The Sep-Pak PS2-Plus cartridges were found more suitable for extracting tricyclazole from water samples than the Supelclean ENVI-18 cartridges. For this cartridge, both methanol and ethyl acetate produced good results. The method was validated with good linearity and with a limit of detection of 0.008 mu gL-1 for a 500-fold concentration through the SPE procedure. The recoveries of the method were stable at 80% and the precision was from 1.1-6.0% within the range of fortified concentrations. The validated method was also applied to measure the concentrations of tricyclazole in real paddy water. 
C18 sorbent; HPLC; Paddy water; Solid phase extraction; Styrene-divinyl benzene copolymer sorbent; Tricyclazole 