Health & Environmental Research Online (HERO)


Print Feedback Export to File
1448109 
Journal Article 
SIMULTANEOUS DETERMINATION OF PARACETAMOL, CAFFEINE, DOMPERIDONE, ERGOTAMINE TARTRATE, PROPYPHENAZONE, AND DROTAVERINE HCL BY HIGH PERFORMANCE LIQUID CHROMATOGRAPHY 
Issa, YM; Hassoun, MEM; Zayed, AG 
2012 
Yes 
Journal of Liquid Chromatography and Related Technologies
ISSN: 1082-6076
EISSN: 1520-572X 
35 
15 
2148-2161 
A high performance liquid chromatographic (HPLC) method was developed for the simultaneous determination of paracetamol, (acetaminophen), caffeine, domperidone, ergotamine tartrate, propyphenazone, and drotaverine HCl. The chromatography was performed in the isocratic mode on a Hypersil C18 BDS column using a mobile phase consisting of 0.02 mole L-1 tetrabutylammonium bisulphate and methanol (100: 45, v/v) at 50 degrees C using UV detection at 210nm. The method was validated by specificity, linearity, limit of detection, and limit of quantification, accuracy, and precision for the aforementioned drugs. The time required for a single analysis amounts to 14min. The response was linear for the drug concentration ranges of 6.20-420.00, 0.80-80.00, 0.80-64.00, 2-40.00, 1.20-240.00, and 3.20-64.00, mu g mL(-1) for PA, CA, DO, ET, PP, and DH, respectively. Accuracy values for the method ranged from 99.40-101.94, 98.48-101.13, 98.97-100.54, 96.47-102.43, 98.92-101.82, and 97.63-100.45% for PA, CA, DO, ET, PP, and DH, respectively. 
caffeine; domperidone; drotaverine HCL; ergotamine tartrate; HPLC; method validation; paracetamol (acetaminophen); propyphenazone 
IRIS
• Methanol (Non-Cancer)
     Search Jan 30 2013
          WOS